THE TRINITY CLOCK

 

 
 

Data

From to [ Shift -1 day | Shift +1 day ] Zoom menu

Plot and as with

Comments (see all | edit)

26 Mar 2018 [16:00]
a few tweaks, with John Aldridge: 1. stop pendulum to lower the prism by a couple of mm to make sure that the twist laser clears the clock-bed beam; 2. cleaned the prism (a fingerprint was causing a big diagonal difraction streak); 3. replace the laser screen with a 1mm polythene sheet; 4. calibrated the tilt level (found that it was pretty good, 1mV/mrad and that the zero-offset knob doesn't introduce scaling errors); 5. restart pendulum and check going, no regulation required.

Download data

Interval between data points: seconds [either 3 seconds (weather data is duplicated) or multiples of 60 seconds (all data is averaged)].

[Raw data: Clock | Weather]


Contact: clockkeeper@trin.cam.ac.uk, Trinity College, Cambridge
// // // // < noscript>
free web stats
//